One: A tilting slope. Can be present at one energy, other energys are OK. This can be corrected by changing the 2T (1T-1R) current. Investigate if seesaw is possible. (minimum tilt of 2% before dose drops)
Second: A step form at the centre of the field. The field is more or less
flat at both sides of this step. This step is caused by a miss
alignment of focus position and flattening filters. Because of the
profile of the photon filters the error for Xhigh will be larger than
for Xlow. Trying to correct this profile using 2T current will give
bad results. If electron flatness is tilted as well look for miss
alignment of focus position. We once found a washer on one side of the
first bending magnet. This washer altered the fringing fields of
this bending magnet moving the focus in the AB plane.
When only photons are effected, investigate the moving and locking
mechanism of the secondary filters. The moving mechanism must have
slack when the filters are locked.
Jan Kok 2010-03-05