It is beneficial to pay attention to good light field alignment. When
Light field setup is OK replacing a bulb will take no more than ten
minutes. Wrong setup results in low light intensity, bad correlation
between light and Xray dimensions, bad field size calibration, moving
cross hairs on isocentre or on the floor.
We use the following procedure with great success.
- Place the bulb in the centre of the optics. (With the power
switched off you can see the filament through the lenses). This
results in maximum light intensity.(non MLC)
- Place the table at about the isocentre height. Fix a rule at the
edge of the table in a way that light from the light field passes
through the hole of the rule. Place a piece of paper on the floor.
Rotate the diaphragm 4 times 90 degrees and mark the projection of
the light spot. Determine the centre of the four marks. Carefully
move the light assembly to match with the centre. Rotate diaphragm
and check if the light spot stays at the same place. If so than the
light focus is on the rotation axis of the diaphragm.
- Check projection of cross hairs at isocentre and re-adjust with
the small mirror (better than 1/2 mm). If OK then check at the floor
(Max 2mm). Check if cross hairs are parallel to blocks (remove
trimmers)
- Ensure if blocks are calibrated. (The easiest and fastest way of
block calibration can be done in a water phantom). Re-position
trimmers to match field size.
- Check magnification factor. Lower table position for example
20cm. Light field dimensions must be 20% larger.
Jan Kok
2010-03-05