Check for offset on one of the four plates. This is a normal error after replacing a chamber. Otherwise check chamber for leaks, cables or SIB board.
After replacing SIB, check for reading temp sensors. When the reading deviates more than 60% change resistor 2K4 to 3K. Otherwise limits cant be set.