According to TSD the focus current must be tuned for minimum Radiation leakage and maximum output. This measurement requires a device that can measure the quotient of output and leakage.
A more simple and fast method is to setup a small field and place a measuring device on the floor. Now tune 1R, 1T, F1, F2 for maximum output (2R 2T at 0 Amp)
Re-tuning Focus current is unusual.
| F1 | F2 | |
| 6 MV | 10 | 12 |
| 8 MV | 10 | 12 |
| 10Mv | 10 | 14 |
| 18MV | 12 | 18 |