One of the few expensive habits of the 25's are the leaking Dosimetry
channels. Especially with the old ``Aluminium'' ionisation chambers
the ladies are (were) frequently forced to clear the added dose before
they could start a treatment. Sometimes leaks are that high that
dosimetry accuracy is influenced. In most cases, these leaks find
their cause in the chamber itselve. This can be determined by
disconnecting the ionisation chamber. By removing the wedge the
ionisation chamber connector can be reached.
(Disconnect connector SK12F1 on the RHCA. Make sure that the cable is ok. )
To get an impression of the leaking rate:
- Start an energy and push the interrupt button (otherwise the
readings of the Dosimetry channels are cleared every five minutes).
- Go to the calibration page.
- Enter in Cal-Item1 the desired Dos Channel. For example 44 for
Dos Chan 1.
- Enter in Cal Part Select 152.
- Note cal part value.
- Wait five minutes and do another readout.
- Subtract both values.
- Max number of count is 142 (for every monitor unit there are 64
counts).
(In the new manual ``Supplementary setup and test
procedures'' chapter 8, a value of 64 counts is given? I don't like
the described procedure as dos. refs are changed and saved.)
When an ionisation chamber starts to leak or gives 300 Volt errors it
may help to knock gently with a small hammer on the side of the
chamber. This trick can expend the life of the ion chamber from 1 hour
to one year. These problems start due to oxidation of the Aluminium
spacers used between the foils. (When the chamber is removed and you
shake it gently you can hear the Aluminium oxide moving through the
chamber).
Leaking can be caused by high humidity. Usually this fault occurs in
the morning after a raining weekend. It can be solved by removing the
wedge and apply hot air for several minutes. Be careful!!
In older machines: the cable tree from the ionisation chamber is also
suspect: disintegrating isolation!
In exceptional cases the DOS PCB are faulty.
For a correct functioning of the 2T, 2R and Hump systems it is
important to check if there is a leak (offset) at one of the measuring
channels at regular intervals. This can be done at the first seconds
after the start button has been depressed, or by opening the fascia
doors.
Jan Kok
2010-03-05